Cereals & Grains Association
Log In

Determining the Fat Acidity of Rough Rice by Near-Infrared Reflectance Spectroscopy

September 1997 Volume 74 Number 5
Pages 556 — 560
Wang-Sheng Li 1 and Jai-Tsung Shaw 1 , 2

PhD candidate and professor, respectively, Department of Agricultural Machinery Engineering, National Taiwan University, 136 Chou-Shan Rd., Taipei 106, Taiwan. Mention of company or trade name is for purpose of description only and does not imply endorsement by the National Taiwan University. Corresponding author. E-mail: m320@ccms.ntu.edu.tw


Go to Article:
Accepted April 25, 1997.
ABSTRACT

Near-infrared reflectance spectroscopy (NIRS) was used to develop calibration curves for determining the fat acidity of whole-kernel and ground rough rice with 13% moisture content at 25°C. Partial-leastsquares regression (PLSR) uses the optimal calibration curve for wholekernel rough rice to measure the coefficient of determination (r2) of validation and standard error of prediction (SEP) of 0.87 and 0.83 mg of KOH/100 g of dry matter, respectively. However, the optimal calibration curve for ground rough rice has a higher r2 of validation and lower SEP of 0.94 and 0.73 mg of KOH/100 g of dry matter, respectively. From 10 to 40°C, the temperature effect causes an increase of 0.24 mg of KOH/100 g of dry matter/°C in the predicted fat acidity of whole-kernel rough rice.



© 1997 by the American Association of Cereal Chemists, Inc.